19. Fast Sweep Time Accuracy Test
ms)
5. Press MARKER
(
PEAK
6. Set
[
FREQUENCY SPAN
)
to 0 Hz,
to 3 MHz,
to 3
MHz, and press TRIGGER
7. Set synthesized sweeper for an amplitude-modulated output.
8. Set function generator controls as follows:
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . triangular wave
AMPLITUDE . . . . . . . . . . . . . .
approximately 1 Vp-p
OFFSET
.
.
.
.
.
.
CAL position (in)
SYM . . . .
. . CAL position (in)
T R I G G E R P H A S E
FREE RUN
MODULATION .
. all out
9. Key in
(
SWEEP TIME
]
5 ms and set function generator for a counter
reading of 2.00
10. Adjust spectrum analyzer TRIGGER LEVEL to place a peak of
the triangular waveform on the first graticule from the left edge
of the CRT display as a reference. (Adjust function generator
amplitude, if necessary, to provide a signal large enough to
produce a stable display). The fifth peak from the reference
should be within
division of the sixth graticule from the left
edge of the display (see Figure 2-34).
11. Using sweep times and function generator frequencies in
2-18, check sweep time accuracy for sweep times
ms by
procedure of step 10.
Figure 2-34. Fast Sweep Time Measurement
ms)
2-64 Performance Tests
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...