6. Impulse and
Resolution
Bandwidth
Switching
Uncertainty
Related Adjustment
Specification
Description
Equipment
Procedure
3 MHz Bandwidth Filter Adjustments
21.4 Bandwidth Filter Adjustments
Down/Up Converter Adjustments
10 Hz bandwidth
30 Hz bandwidth
100 Hz to 1 MHz bandwidth
1 .O
3 MHz bandwidth
30
and 100
bandwidth switching uncertainty figures only
applicable
Relative Humidity.
The CAL OUTPUT signal is applied to the input of the spectrum
analyzer. The deviation in peak amplitude of the signal trace is then
measured as each resolution bandwidth filter is switched in.
None required
1.
2.
3.
4.
5.
Press
Connect CAL OUTPUT to SIGNAL INPUT 2.
Key in the following control settings:
(
C
E
NT
E
R
FREQUENCY]
S P A N
) .
MHz
5 MHz
. - 8
1 MHz
Press LOG
[
ENTER
and key in 1
Press MARKER
(
PEAK
SEARCH)
Key in settings according to
4-6. Press MARKER
[
PEAK SEARCH
)
at each setting, then read the amplitude deviation
from the MARKER
readout at the upper right of the display
(see Figure
allowable deviation for each resolution
bandwidth setting is shown in the table.
Option 462
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...