18. Frequency Control Adjustments
F R E O U E N C Y C O N T R O L
S T A R T
F M S P A N
S T O P
T I L T
Y T O
R E F
R E F
F A S T
Figure 3-71. Location of Frequency Control Adjustments
5. Connect DVM to
and ground to
6. Adjust
TUNE REF
for DVM indication of -10.285
V dc. See Figure 3-71 for location of adjustment.
7.
Key in
[
CENTER FREQUENCY
)
10 MHz,
SPAN
]
0 Hz, Trace
A
[
CLEAR
-
WRITE
),
Sweep
Scale LIN.
Start-Up Time
8.
Key in
left
TIME)
Marker
Adjust marker to the
edge of the CRT. Key in (SHIFT)
then key in (SHIFT)
three times. CRT annotation should indicate SWEEP
GEN measured sweep time.
1 second start-up time:
Note
The start-up time measured in step 8 uses the
function that displays a sweep time value which is 1% to 5%
longer than the actual spectrum analyzer sweeptime. This error is
compensated when using the shift F function to adjust the sweep
times in the following procedure.
9. Key in Marker
then
Slow Sweep
10. Key in [SHIFT)
three times and note the CRT
Adjustment
annotation. The annotation should indicate SWEEP GEN
measured sweep time of (1 .OO + start-up time from step 8)
s. To adjust sweep time, adjust
SLOW slightly, then
key in
and note new SWEEP GEN measured
sweep time as indicated by CRT annotation. Repeat this process
until the 1 sweep time is within
Note
Adjusting
CW decreases the sweeptime.
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...