10. Step Gain and 18.4 MHz Local Oscillator Adjustments
1
Gain Step Checks
22. Key in
[
REFERENCE LEVEL
)
-19.9
Set step attenuators to 15
Press MARKER
twice to establish a new reference.
23. Key in
(
REFERENCE LEVEL
)
-17.9
Set step attenuators to 13
24. MKR A level, as indicated by CRT annotation, should be .OO
If not, increase or decrease the value of
Refer to
3-3 for range of values.
25. Key in
(
REFERENCE LEVEL
)
-15.9
Set step attenuators to 11
26. MKR A level should be .OO
If not, increase or decrease
the value of
Refer to Table 3-3 for range of values.
27. Key in
[
REFERENCE LEVEL
)
-11.9
Set step attenuators to 7
28. MKR A level should be .OO
If not, increase or decrease
the value of
Refer to
3-3 for range of values.
Gain Step Adjustment
29. Key in LIN, (SHIFT) (AUTO] (resolution bandwidth), and
(
REFERENCE LEVEL
)
-19.9
Set step attenuators to 13
Press
MARKER
twice to establish a new reference.
30. Key in
[
REFERENCE LEVEL
)
-18.0
Set step attenuators to 11
31. Adjust
VR for MKR A level of + 0.10
See Figure 3-49
for location of adjustment.
32. Remove all test equipment from the spectrum analyzer. Connect
CAL OUTPUT to RF INPUT.
18.4 MHz Local Oscillator Adjustment
33. Press
PRESET
]
and
34. Set front-panel FREQ ZERO control to midrange.
35. Adjust
FREQ ZERO to peak signal trace on CRT. See
Figure 3-49 for location of adjustment.
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...