23. Analog-To-Digital Converter Adjustments
Procedure
1. Position instrument upright as shown in Figure 3-85 and remove
top cover.
2. Set LINE switch to ON and press
PRESET
).
Standard Procedure
3. Procedure using Low-Noise DC Supply is illustrated in Figure 3-93.
a. Key in
TRACE A and SWEEP
b. Disconnect cable 0 (black) from sweep ramp input
c. Short
to
or connect SMB snap-on short to
1.
d. Connect the oscilloscope’s
probe to
and ground
the probe’s ground to the
section’s card cage.
e. Set the oscilloscope settings as follows:
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . 0.1
time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
f. Adjust
OFFS for a square wave displayed on the
oscilloscope. The square wave should be approximately 4 V
p-p
.
See Figure 3-86 for location of adjustment.
A N A L O G - T O - D I G I T A L C O N V E R T E R
( B e n e a t h C o v e r )
G A I N O F F S
Figure 3-86.
Location of Analog-To-Digital Converter Adjustments
g. Remove short from
and
or disconnect the
SMB snap-on short from
h. Press
PRESET
).
i. Press MARKER
1498 (MHz), and
j. Connect DVM to
and ground to
Set DVM
for V
k. Connect output of the Low-Noise DC Supply to
Adjust
the Low-Noise DC Supply for DVM indication of + 10.000
dc.
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...