3. Sweep Time Accuracy Test
ms)
3. Sweep Time
Accuracy Test
Related Adjustment
Frequency Control Adjustments
Specification
for sweep times
seconds
for sweep times
seconds
Description
Preferred Procedure
This test is for sweep times
ms. For faster sweep times, refer to
Fast Sweep Time Accuracy Test (Test 19).
A universal counter is connected to the
RECORDER
OUTPUT (on
the
rear panel) of the spectrum analyzer. The counter is
used in time interval mode to determine the “pen down” (sweep time)
interval of the
RECORDER OUTPUT. The
output
voltage level corresponds directly to the sweeping of the analyzer
(pen down = OV) and not-sweeping of
the
analyzer (pen up =
A
DVM is used to set the appropriate trigger level for the counter.
Alternate Procedure
Perform this procedure if the equipment for
the
preferred procedure
is unavailable.
Sweep time accuracy for sweep times
ms can also be measured
using the HP
internal frequency counter for a time interval
measurement.
SPECTRUM ANALYZER
UNIVERSAL
COUNTER
DIGITAL
VOLTMETER
Figure 2-4. Sweep Time Accuracy Test Setup
Performance Tests 2-9
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...