7 .
8.
9 .
10.
5. 6
Resolution Bandwidth Selectivity
Vary spectrum analyzer settings according to Table 4-4. Press
SWEEP
and measure the 60
bandwidth for each
resolution bandwidth setting by the procedure of steps 4 through
6. Record the value in
4-4.
Record the 6
bandwidths from Table
in
4-4.
Calculate the bandwidth selectivity for each setting by dividing
the 60
bandwidth by the 6
bandwidth. The bandwidth
ratios should be less than the maximum values shown in
4-4.
The 60
bandwidth for the 10 Hz resolution bandwidth setting
should
be
less than 100 Hz.
Figure 4-4. 60
Bandwidth Measurement
Option 462