O p t i o n 4 0 0
2.
3. Press
4. Key in the following analyzer settings:
5.
6.
7.
17. Line-Related Sidebands
Figure 2-31. Line Related Sidebands Measurement
Press
PRESET
].
Connect CAL OUTPUT to SIGNAL INPUT 2.
Press
8 and adjust AMPTD CAL for a MARKER amplitude
of -10. 00
l
FREQUENCY] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20 MHz
L E V E L
] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
- 1 0
(FREQU
ENCY
S
PAN
] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3
Wait for completion of the sweep, then press MARKER
[
P E A K
Press (SHIFT)
SWEEP
10
to initiate
video averaging of 10 sweeps. Wait for completion of sweeps.
Press MARKER
and position movable marker at
the peak of each line related sideband (400 Hz, 800 Hz, and 1200
Hz). The MARKER
A
amplitude for each sideband should be
400 Hz
800 Hz
1200 Hz
Performance Tests 2-61