8. 21.4 MHz Bandwidth Filter Adjustments
22. Adjust
CTR to center signal on center graticule line.
Adjust
SYM for best symmetry of signal. See Figure 3-42
for location of adjustments. If unable to adjust
SYM
for satisfactory signal symmetry, increase or decrease value of
Refer to
3-3 for range of values.
23. All crystal filter bypass networks are removed. Signal should be
centered and symmetrical. If not, go back to step 16 and repeat
adjustments.
24. Press MARKER
SEARCH
]
and MARKER In].
25. Key in
SPAN
]
20
3
and MARKER
SEARCH].
26. Adjust
XTAL to align markers on display. MARKER
A level should indicate 1.00 X. See Figure 3-42 for location of
adjustment.
LC Adjustments
27. Disconnect cable 97 (white/violet) from
and reconnect to
Reconnect cable 89 (gray/white) to
28. Key in
100
and
[
FREQUENCY SPAN
]
200
29. Press
[
REFERENCE LEVEL
)
and adjust DATA knob to place signal peak
two division from the top graticule line.
30. Adjust
LC CTR and
LC CTR for maximum
MARKER level as indicated by CRT annotation. See Figure 3-43
for location of adjustments. If unable to adjust
and
LC CTR adjustments for satisfactory signal amplitude,
increase or decrease value of
and
Refer to
3-3 for range of values.
A t t e n u a t o r
F i l t e r
Figure 3-43.
Location of
21.4 MHz LC Filter and Attenuation
Adjustments
31. Key in
1 MHz and
1 MHz.
32. Press MARKER
(
PEAK SEARCH
]
and MARKER
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...