5. Resolution Bandwidth Selectivity Test
6. Read the 60
bandwidth for the 3 MHz resolution bandwidth
setting from the MARKER A frequency readout (see Figure 2-7)
and record the value in
2-9.
7. Vary spectrum analyzer settings according to Table 2-9. Measure
the 60
bandwidth for each resolution bandwidth setting by the
procedure of steps 4 through 6 and record the value in
2-9.
8. Record the 3
bandwidths from
2-8 in
2-9.
9. Calculate the bandwidth selectivity for each setting by dividing
the 60
bandwidth by the 3
bandwidth. The bandwidth
ratios should be less than the maximum values shown in
2-9.
10. The 60
bandwidth for the 10 Hz resolution bandwidth setting
should be less than 100 Hz.
Figure 2-7. 60
Bandwidth Measurement
2-16 Performance Tests