13. 20 MHz Reference Adjustments
Procedure
1. Position instrument on right side as shown in Figure 3-56 and
remove bottom cover. Remove
20 MHz Reference and install
on extenders. See Figure 3-57 for the location of
components.
2. Set LINE switch to ON and press
PRESET
).
3. Set rear-panel FREQ REFERENCE INT/EXT switch to INT.
Disconnect cable 2 (red) from
Connect power meter to
output of Time Base
using cable 2 (red). Note power
meter indication for reference later.
4. Reconnect
Time Base output to
5.
to Ground. Set the HP
Spectrum
Analyzer to
[
CENTER FREQUENCY
)
20
MHz,
SPAN
)
1 MHz,
LEVEL
)
and
100
Connect
to RF INPUT of HP
Analyzer and set
[
REFERE
N
CE
to place of signal at reference line (top graticule
line).
6. Set HP
Spectrum Analyzer to 1 dB/division SCALE and
reset reference level to place peak of signal at reference line.
7. Connect DVM to
and ground to
Adjust
COMB DRIVE
for DVM indication of + 0.1 V dc.
Disconnect DVM. (If DVM remains connected, it may load circuit.)
See Figure 3-57 for location of adjustment.
R E F E R E N C E
Figure 3-57. Location of 20 MHz Reference Adjustments
8. Adjust
DOUBLER
to lower signal peak approximately
3
Adjust
CENTER FREQ
to peak signal on HP
Spectrum Analyzer display. Next, adjust
DOUBLER
for signal peak.
9. Disconnect cable 2 (red) from
and connect
OUTPUT of
frequency synthesizer to
Set FREQUENCY of frequency
synthesizer to 10.17 MHz and set AMPLITUDE to + 3
Set HP
3-104 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...