13. 20 MHz Reference Adjustments
Press
Press
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on
Vmarker 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
start marker . . . . . . . . . . . . . . . . . place at
crossing
stop marker . . . . . . . . . . . . . place at next
crossing
2 1. Oscilloscope display should be a 10 MHz signal of TTL level; less
than
to greater than
See Figure 3-58 for a typical
signal.
- 6 0 . 0 0 0
4 0 . 0 0 0
1 4 0 . 0 0 0
2 0 . 0
Figure 3-58. Typical Signal at
22. Install
20 MHz Reference without extenders and reconnect
cable 7 (violet) to
23. Connect power meter to front-panel CAL OUTPUT.
24. Adjust
CAL LEVEL
for power meter indication of
-10.0
25. the
Comb Generator must be readjusted after adjusting the
20 MHz Reference. Refer to Adjustments 22, Comb Generator
Adjustments, for adjustment procedure.