2. High-Voltage Adjustment (SN
and Below)
Warning
With the protective cover removed in the following step, do not
place hands near the
High-Voltage assembly. High voltage
(approximately -4000 V dc) can be present even when the ac line
cord is disconnected.
10. Wait at least one minute for capacitors to discharge to a safe
level.
11. Remove
the
protective cover from the
High-Voltage
Regulator. A label should be visible on the
High-Voltage
Transformer. Record the voltage listed on the label for use in step
15.
Note
If the label is missing, use the nominal value of -3790 V dc.
12. Connect the high-voltage probe to
See Figure 3-7 for
the location of the test point.
Warning
With power supplied to the instrument,
is at a voltage
level of approximately -4000 V dc. Be extremely careful.
V o l t a g e
R e g u
Figure 3-7. Location of Label and Test Point
13. Reconnect ac line cords to both instrument sections. Set the LINE
switch to ON.
14. Wait approximately 30 seconds for the dc regulator circuits to
stabilize.
15. Adjust
HV ADJ for a DVM indication equal to the
calibration factor (calculated in step 8) times the voltage labeled
on the top of
High-Voltage Regulator (noted in step 11).
See
Figure 3-6 for the location of the adjustment.
V dc
E
X A M P L E
:
3-32 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...