16. Second IF Amplifier and Third Converter Adjustment
Note
Place the Markers as accurately as possible within the cursor markers
for maximum frequency accuracy.
Note
Third Converter
Adjustment
18. On the synthesized sweeper, press
should read
between 7 and 14 MHz.
19. On the synthesized sweeper, press
OFF and
OFF.
20. Set the synthesized sweeper FREQUENCY MARKERS
to 251.4
MHz and
to 351.4 MHz.
21. Set the Scalar Network Analyzer
to 10
and set
(REF LEVEL) to + 14
Set the REF POSN to one division down
from the top.
22. Adjust
for minimum amplitude response at 258.4 MHz.
Refer to Figure 3-64 for the location of the
adjustments.
Refer to Figure 3-66 for the typical response when the
filter is properly adjusted.
2 8 0
Figure 3-66. Minimum Image Response at 258.4 MHz
23. Repeat the adjustments in steps 13 and 22 to assure that the
is between 7 MHz and 14 MHz and the image response
at 258.4 MHz is minimized.
Remember to use the appropriate set up for steps 13 and 20.
24.
25.
26.
27.
28.
Disconnect the cables from
and
and reconnect the
instrument cables.
Disconnect cable 83 (gray/orange) from
and connect
to the input of HP
Spectrum Analyzer, using a BNC to
SMB test cable.
Press
PRESET
)
on the HP
Spectrum Analyzer, then
key in
[
CENTER FREQUENCY
)
280
MHz,
[
FREQUENCY SPAN
)
2
MHz. Set
MARKER
+ 2
and
[
ENTER
1
Adjust
AMPTD
for maximum signal level as indicated
on spectrum analyzer display.
Disconnect spectrum analyzer and reconnect cable 83
(gray/orange) to
Adjustments 3-115
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...