8. 21.4 MHz Bandwidth Filter Adjustments
33. Key in
100
200
and MARKER
[
PEAK
SEARCH].
34. Adjust
LC to align makers on display. MARKER A level
should indicate 1.00 X. See Figure 3-43 for location of adjustment.
35. Repeat steps 31 through 34 until no further adjustment is
necessary.
Adjustments
36. Key in
30
[
FREQUENCY SPAN
]
100
Press
MARKER
37. Connect crystal filter bypass network between
and
38. Press
(
REFERENCE LEVEL
)
and adjust DATA knob to place signal peak
two division from the top graticule line.
39. Adjust
CTR to center signal on center graticule line.
Adjust
SYM for best symmetry of signal. See Figure 3-44
for location of adjustments. If unable to adjust
SYM
for satisfactory signal symmetry, increase or decrease value of
Refer to Table 3-3 for range of values.
A t t e n u a t o r - B a n d w i d t h F i l t e r
Figure 3-44.
Location of
21.4 MHz Crystal Filter Adjustments
40. Remove crystal filter bypass network from between
and
41. Adjust
CTR to center signal on center graticule line.
Adjust
SYM for best symmetry of signal. See Figure 3-44
for location of adjustments. If unable to adjust
SYM
for satisfactory signal symmetry, increase or decrease value of
Refer to
3-3 for range of values.
42. Press MARKER
SEARCH
)
and MARKER
43. Key in
[
FREQUENCY SPAN
]
10
44. Key in
3
and MARKER
SEARCH
).
Adjustments 3-81
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...