3. Sweep Time Accuracy
ms)
Equipment
Universal Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Digital Voltmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Procedure
Sweep Times
ms
1. Connect equipment as shown in Figure 2-4.
2. Press
PRESET
)
on the spectrum analyzer.
3. Key in the following settings:
(
C E N T E R F R E Q U E N C Y
) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
500
[
F R E Q U E N C Y S P A N
]
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0
4. Set up the universal counter as follows:
a. Set all front panel keys in “out” position.
b. Set POWER switch to ON.
c. Set GATE TIME vernier control to 9 o’clock.
d. Set
A switch to COM A position.
e. Depress T.I. A B switch (making sure the blue shift key is
out).
f. Set Channel A trigger level to trigger on negative slope.
g. Set Channel B trigger level to trigger on positive slope.
h. Set both Channel A and Channel B
switches to
i. Connect the digital voltmeter to Channel A TRIGGER LEVEL
OUT. (Be sure to ground the DVM properly.)
j. Adjust Channel A trigger level to set a DVM voltage reading of
0.3 v.
k. Repeat steps i and j for Channel B.
5. Set analyzer
TIME
)
to 20 ms. Allow the universal counter
enough time to settle at this sweep time.
Performance Tests
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...