
2. High-Voltage Adjustment (SN
and Below)
Note
The accuracy of the high-voltage probe is specified for a probe
connected to a dc voltmeter with 10
input resistance. HP
and HP
digital voltmeters have a 10
input resistance on the
100 V and 1000 V ranges. All measurements in this procedure should
be performed with the DVM manually set to the 100 V range (fOO.OOO
on the HP
display).
H I G H
V O L T A G E
R E G U L A T O R
,
.
Figure 3-6. Location of High Voltage Adjustments
4. Set the LINE switch to ON. Set
the
front-panel INTENSITY
control fully counterclockwise (CRT beam at cut-off) to prevent
possible damage to the CRT.
5. Note the DVM indication at
DVM Indication:
6. Connect the high-voltage probe to the DVM. Connect the probe to
7. Note the DVM indication.
DVM Indication:
8. Divide the DVM indication in step 7 by the DVM indication in
step 5. This gives the calibration factor needed to compensate for
high-voltage probe error.
Calibration Factor:
9. Disconnect the high-voltage probe from
Set the
LINE switch to STANDBY. Remove the ac line cord from both
instrument sections.
Warning
The MAINS power-on indicator
(red LED) should
be completely off before proceeding with this procedure. See
Figure 3-6. The indicator will remain lit for several seconds after
the ac line cord has been removed, and will go out slowly (the
light becomes dimmer until it is completely out).
Adjustments 3-3 1
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...