20. 50
MHz
Voltage-Tuned
Oscillator
Adjustments
Reference
Related Performance
Test
Description
Equipment
Procedure
RF Section:
All 50 MHz Voltage-Tuned Oscillator (VTO)
Frequency Span Accuracy Test
Center Frequency Readout Accuracy Test
First, the voltage reference for the Shaping network is set by
measuring the voltage required to tune the 50 MHz Oscillator to its
high limit (57.5 MHz) and then setting the reference voltage (+ 15 VR)
to that voltage.
Next, the
tuning accuracy is adjusted at both the low and high
end by setting the tune voltage to the proper levels to tune the
to its low and high end limits (42.5 MHz and 57.5 MHz). This is done
using the output of the tuning DACS from the
Frequency Control;
therefore, it is necessary that the DAC adjustments on the Frequency
Control have been performed before adjusting the 50 MHz VTO.
SPECTRUM ANALYZER
DIGITAL VOLTMETER
Figure 3-78. 50 MHz Voltage-Tuned Oscillator Adjustments Setup
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . HP
1. Position Instrument on right side as shown in Figure 3-78 and
remove bottom cover. Remove All 50 MHz Voltage-Tuned
Oscillator and place on extenders.
2. Set LINE switch to ON and press
PRESET
).
3-130 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...