18. Frequency Control Adjustments
18. Frequency
Control
Adjustments
Reference
Related Performance
Tests
Description
Equipment
Procedure
RF Section:
Frequency Control
Sweep Time Accuracy Test
Frequency Span Accuracy Test
Center Frequency Readout Accuracy Test
The sweep reference voltage is adjusted and then the sweep times are
adjusted for proper tolerances. The sweep tune voltage is adjusted.
Then the YTO DAC,
DAC, and LSD
DAC are adjusted, each
to within its tolerance. Next,
the
Start and Stop frequencies are
adjusted. FM Span is adjusted next for the proper amount of FM
deviation.
SPECTRUM ANALYZER
DIGITAL VOLTMETER
FREPUENCY COUNTER
Figure 3-70. Frequency Control Adjustments Setup
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . HP
Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
1. Place instrument
on
right side with IF-Display facing right as
shown in Figure 3-70 and remove bottom cover.
2. Set LINE switch to ON and press
PRESET
).
3. Connect DVM to
and ground to
4. Adjust
REF
for DVM indication of 10.00
V
dc.
See Figure 3-71 for location of adjustment.
Adjustments
19
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...