12. Amplitude Fidelity
12. Amplitude
Fidelity Test
(For instruments with Option 857, refer to Chapter 5.)
Related Adjustment
Log Amplifier Adjustments
Specification
Log:
Incremental
over 0 to 80
display
Cumulative
3 MHz to 30 Hz Resolution Bandwidth
max over 0 to 80
display (20
1.5
max over 0 to 90
display
Linear:
of Reference Level for top
divisions of display
Description
Amplitude fidelity in log and linear modes is tested by decreasing the
signal level to the spectrum analyzer in 10
steps with a calibrated
signal source and measuring the displayed amplitude change with the
analyzer’s MARKER A function.
f
Figure 2-21. Amplitude Fidelity Test Setup
Performance Tests