14. Residual Responses Test
9.
10.
11.
Figure 2-24. Residual Responses Measurement
Press SWEEP
and wait for completion of sweep. Look for
any residual responses at or above the display line. If a residual
is suspected, press SWEEP
again and see if the response
persists. A residual will persist on repeated sweeps, but a noise
peak will not. Any residual responses must be
Option 400:
Any residual 500 Hz to 2.5
must be
any residuals
must be
If a response appears marginal, do the following to determine
whether or not it exceeds the specification.
a. Press
b. Press MARKER
and place the marker on the peak of
the response in question.
c. Press MARKER
then activate SWEEP
d. Reduce
[
FREQUENCY SPAN
]
to 1 MHz or less. The amplitude of
the response should be
(below the display line).
e. Press (RECALL]
to resume the search for residuals.
Step
[
CENTER FREQUENCY
)
to 1510 MHz with
checking for
residual responses at each step by the procedure of steps 9 and
10. There should be no residual responses at or above the display
line below 1500 MHz.
Maximum Residual Response
MHz
2-46 Performance Tests
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...