19. Second Converter Adjustments
Second Converter
53. Repeat steps 14 through 19 to ensure that Second LO frequency
Final Adjustments
and shift are still properly adjusted.
54. Check the
at the 3
points for both the
LO
and On the scalar network analyzer, press
Max. Press
cursor A
and set the cursor at the -3
point
Press
cursor A, cursor A, and set the cursor to the corresponding -3
point on the opposite side of the signal. The cursor should now
read 0
55. On the synthesized sweeper, press
and place the marker on
either cursor A. Press
and place the marker on the cursor A
on the opposite side of the trace.
56. On the synthesized sweeper, press
and read the
shown on the ENTRY DISPLAY. Press
OFF. See
Figure 3-74 and Figure 3-75.
57. Disconnect all test equipment from HP
Spectrum Analyzer
and reconnect all cables within the instrument: cable 80
(gray/black) between
and
and cable 92 (white/red)
between
and
58. Connect HP
Spectrum Analyzer CAL OUTPUT to SIGNAL
INPUT 2. Key in
(
CENTER FREQUENCY
)
20 MHz,
(
FREQUENCY SPAN
)
1
MHz,
(
REFERENCE
-7
SCALE LOG
[
ENTER
1
300
59. Key in
, (
PEAK SEARCH
]
Key in
and note
signal amplitude as indicated by marker level CRT annotation.
60. Continue to key in
then (SHIFT)
while adjusting
for maximum amplitude and the same amplitude in both
states of the Second LO
61. Reinstall RF Converter in instrument. See installation procedure
in RF Section of Troubleshooting and Repair Manual, Volume 1.
Adjustments 3-129
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...