18. Calibrator
Amplitude
Accuracy
Related Adjustment
20 MHz Reference Adjustments
Specification
-10
Description
The output level of the calibrator signal is measured with a power
meter.
Figure 2-32. Calibrator Amplitude Accuracy
Setup
Equipment
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Adapter, Type N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . HP 1250-0077
Procedure
1. Connect equipment as shown in Figure 2-32.
2. Measure output level of the CAL OUTPUT signal. The value should
be -10.0
2-62 Performance Tests