15. Spurious Responses Test
signal for a marker indication of -20.00
(-30.0
at the
input mixer with 10
of input attenuation).
6 On the spectrum analyzer, key in MARKER
FREQUENCY
)
560 MHz, MARKER
SEARCH
)
to position
a second marker on the peak of the second harmonic distortion
product of the 280 MHz input signal. The response should be
below the display line
below the input signal level).
Second Harmonic
7. On the synthesized sweeper, key in
LEVEL
)
decrease
the amplitude of the 280 MHz signal by 10
8. On the spectrum analyzer, key in MARKER
FREQUENCY
)
280 MHz,
[
REFERENCE LEVEL
)
-30
DISPLAY LINE [ENTER] -105
MARKER
SEARCH
)
to
position a marker on the peak of the displayed 280 MHz signal.
9. On the synthesized sweeper, press
[
POWER LEVEL
]
and use the
ENTRY knob to adjust the amplitude of the displayed 280 MHz
signal for a marker indication of -30.00
(-40.0
at the
input mixer with 10
of input attenuation).
10. On the spectrum analyzer, key in MARKER
[
CENTER FREQUENCY
)
840 MHz, MARKER
[
PEAK SEARCH
)
to position
a second marker
on
the peak of the third harmonic distortion
product of
the
280 MHz input signal. The response should be
below the display line
below
the
input signal level).
Third Harmonic
Performance Tests 2-51
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...