8. 21.4 MHz Bandwidth Filter Adjustments
Equipment
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . .HP
10
Step Attenuator . . . . . . . . . . . . . . HP
Option
1
Step Attenuator . . . . . . . . . . . . . . . HP
Option
Crystal Filter Bypass Network (2 required) . . . . Refer to Figure 3-91
Procedure
1. Position instrument upright as shown in Figure 3-40 and remove
top cover.
2. Set LINE switch to ON and press
PRESET
].
+ 10 V Temperature Compensation Supply Check
3. Connect DVM to
(+ 10 VF).
4. DVM indication should be between
V dc and + 10.0 V dc. If
voltage is within tolerance, proceed to next step. If voltage is not
within tolerance, refer to Adjustment 10, Step Gain and 18.4 MHz
Local Oscillator Adjustments, for adjustment procedure.
LC Adjustments
5. Set step attenuators to 0
6. Disconnect cable 97 (white/violet) from
and connect to
1.
7.
Key in
20
MHz,
100
[
FREQUENCY SPAN
)
200
and press LIN pushbutton.
8.
Press
(
REFERENCE LEVEL
)
and adjust front-panel knob to set signal
peak on screen two divisions from the top graticule.
9. Adjust
LC CTR and
LC CTR for maximum
MARKER level as indicated by CRT annotation. See Figure
for location of adjustments. If unable to adjust LC CTR
adjustments for satisfactory signal amplitude, increase or decrease
value of
and
Refer to
3-3 for range of
values.
\
Figure 3-41.
Location of
21.4 MHz LC Filter Adjustments
3-78 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...