23. Analog-To-Digital Converter Adjustments
1. Adjust
GAIN for a square wave displayed on the
oscilloscope. The square wave should be approximately 4
See Figure 3-86 for location of adjustment.
Alternate Procedure
4. Procedure without using Low-Noise DC Supply:
a. Press
b. Key in TRACE A
and SWEEP (SINGLE).
c. Disconnect cable 0 (black) from sweep ramp input
d. Short
to
or connect SMB snap-on short to
e. Connect DVM to
and ground to
Set DVM
for V ac.
f. Adjust
OFFS until
the
level at
is at a
maximum ac voltage as indicated by the DVM (approximately
2.0 V ac). See Figure 3-86 for location of adjustment.
g. Remove short from
and
Reconnect cable 0
(black) to
1.
h. Press
PRESET
].
i. Connect DVM to
and ground to
Set DVM
for V dc.
j. Press SWEEP
Note DVM reading at end of the sweep.
The voltage will begin to drift immediately after the sweep
ends. Therefore, the first indication after the sweep ends is the
valid indication. It may be helpful to press
several times
to ensure a valid indication at the end of the sweep.
k. If DVM indication is 10.020
V dc at
the
end of the
sweep, no further adjustment is necessary. Otherwise, adjust
GAIN and repeat step j until the voltage at the end of
the sweep is + 10.020
V dc.
Adjustments 3-141
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...