16. Second IF
Amplifier and
Third Converter
Adjustment
Reference
RF Section:
Second IF Amplifier
Third Converter
Description
A synthesized sweeper is used to inject a signal of 301.4 MHz at -20
in to
the
Second IF Amplifier. The output of the amplifier is
displayed on a scalar network analyzer. The amplifier is adjusted for
a
of greater than 7 MHz and less than 14 MHz centered at
301.4 MHz. Its gain should be greater than 14
and less than 17
A spectrum analyzer is used to view the output of the 280 MHz
Oscillator on
the
Third Converter and the oscillator is centered in
its adjustment range.
SYNTHESIZED WEEPER
ANALYZE
METER
2 0
ATTENUATOR
NETWORK ANALYZER
SPECTRUM ANALYZER
U n d e r
Figure 3-63. Second IF Amplifier Adjustments Setup
Equipment
Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . .
HP
Synthesized Sweeper . . . . . . . . . . . . . . . . . . . . . . . .
HP
Scalar Network analyzer . . . . . . . . . . . . . . . . . . . . . . . HP
Power Splitter . . . . . . . . . . . . . . . . . . . . . .
HP
Opt. 001
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Detector (2 required) . . . . . . . . . . . . . . . . . . . . . . . . .
HP
20
Attenuator . . . . . . . . . . . . . . . . . . . . HP
Opt. 020
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...