19. Fast Sweep Time Accuracy Test
ms)
19. Fast Sweep
Time Accuracy
Test
ms)
Related Adjustment
None
Specification
for sweep times
seconds
Description
The triangular wave output of a function generator is used to
modulate a 500 MHz signal which is applied to the spectrum analyzer
SIGNAL INPUT. The signal is demodulated in the zero span mode to
display the triangular waveform. Sweep time accuracy for sweep
times
ms is tested by checking the spacing of the signal peaks on
the displayed waveform.
SPECTRUM ANALYZER
SYNTHESIZED SWEEPER
Figure 2-33.
Sweep Time Accuracy
ms
Setup)
Equipment
Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Universal Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Signal Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Procedure
1. Connect equipment as shown in Figure 2-33.
2. Press
on spectrum analyzer.
3. Key in analyzer settings as follows:
F R E Q U E N C Y
)
..................................... 500
MHZ
(
F R E Q U E N C Y S P A N
) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 0 0
4. Set synthesized sweeper for an output frequency of 500 MHz and
an output power level of -10
Performance Tests 2-63
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...