16. Residual FM Test
Figure 2-28. Bandwidth Filter Slope Measurement
8. Compute the detection slope of the 30 Hz filter between the
markers by dividing the MARKER A amplitude by the MARKER A
frequency:
filter slope = MARKER A amplitude/MARKER Afrequency =
9. Press SWEEP
10. Change
SPAN
]
to 0 Hz. Readjust FREQ ZERO, if
necessary, to position the trace at the center graticule line. The
amplitude variations of the trace (see Figure 2-29) represent the
analyzer residual FM.
Performance Tests
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...