2. High-Voltage Adjustment (SN
and Below)
DIGITIZING
OSCILLOSCOPE
HI-VOLTAGE
SIGNAL ANALYZER
Figure 3-5. High Voltage Adjustment Setup
Equipment
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DC High-Voltage Probe (1000: 1 divider) . . . . . . . . . . . . . . . . . . HP
Display Adjustment PC Board (service accessory) . . . . . .
Digitizing Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Divider Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Function Generator
required) . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
High-Voltage
Adjustment Procedure
Warning
In the following procedure, it is necessary to probe voltages
which, if contacted, could cause serious personal injury. Use
a nonmetallic alignment tool when making adjustments. Be
extremely careful.
Warning
Do not attempt to measure the CRT filament voltage directly. The
filament voltage is referenced to the high-voltage cathode and
can only be measured safely with a special high-voltage true-rms
voltmeter and probe.
1. Set the spectrum analyzer’s LINE switch to STANDBY.
2. Remove the top cover from the IF-Display Section, and
connect
the equipment as shown in Figure 3-5 and described in the
following steps.
3. Set the DVM to
the
100 V range, and connect the DVM to
(+ 100 V). Do not use the high-voltage probe. See
Figure 3-6 for
the
location of
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...