7. 3 MHz Bandwidth Filter Adjustments
10. Adjust
CTR for minimum amplitude of signal peak.
Adjust
SYM for best symmetry. Repeat adjustments to
ensure that the signal is
and adjusted for best symmetry.
See Figure 3-38 for location of adjustments.
11. Remove crystal filter bypass network near
SYM.
12. Adjust
CTR for minimum amplitude of signal peak.
Adjust
SYM for best symmetry of signal. Repeat
adjustments to ensure that signal is
and adjusted for best
symmetry. See Figure 3-38 for location of adjustments.
13. Remove crystal filter bypass network near
SYM.
14. Adjust
CTR for minimum amplitude of signal peak.
Adjust
SYM for best symmetry of signal. Repeat
adjustments to ensure that signal is
and adjusted for best
symmetry. See Figure 3-38 for location of adjustments.
15. Remove crystal filter bypass network near
SYM.
16. Adjust
CTR for minimum amplitude of signal peak.
Adjust
SYM for best symmetry of signal. Repeat
adjustments to ensure that the signal is
and adjusted for
best symmetry. See Figure 3-38 for location of adjustments.
17. Signal should be centered on center graticule line on CRT display.
If signal is not centered, go back to step 3 and repeat adjustments
of each filter stage.
Filter Peak Adjust
18. Press
PRESET
].
19. Key in
TIME
]
20 ms,
[
FREQUENCY SPAN
)
0 Hz,
10 Hz,
[
R E F E R E N C E L E V E L
)
-20
20. Set the frequency synthesizer for 21.400 MHz at an amplitude
level of -25.0
21. Disconnect cable 97 (white/violet) from
and connect
output of the frequency synthesizer to
using BNC to SMB
snap-on cable.
22. Set the oscilloscope following settings:
Channel 1
amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.005
time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...0.2
mag x
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(vertical)
.
coupling
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..ac
probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Channel 2
amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
coupling
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ac
probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
23. Connect oscilloscope Channel 1 probe to
(left side of
SYM) and Channel B probe to
(left side of
SYM).
24. Adjust frequency synthesizer output frequency to peak Channel 1
display.
3-74 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...