16. Residual FM Test
13. Press MARKER
and position movable marker at the lowest
point on
the
trace (see Figure 2-30). Read the MARKER A
amplitude from the display and record its absolute value.
MARKER A amplitude = p-p amplitude =
14. Divide the peak-to-peak amplitude by the slope computed in step
8 to obtain the residual FM:
p-p amplitude/filter slope = residual FM
=
Hz
= residual FM
The residual FM should be less than 3 Hz.
15. Press
16. Press
9 and adjust FREQ ZERO for a maximum amplitude
trace.
Performance Tests 2-59