8. Frequency Response
25. After completion of sweep, press DISPLAY LINE
on the
spectrum analyzer. Use the Display Line to measure the maximum
and minimum points on
the
trace. Record the measurements
below.
SIGNAL INPUT 1
(100
to 20 MHz)
Maximum
Minimum
26. Measure and record signal level at start of trace (100
SIGNAL INPUT 1
(100
27. Connect output of frequency synthesizer to SIGNAL INPUT 2.
Activate this input with the pushbutton.
Option 001.
Do not use HP
Minimum Loss Pad. Set
frequency synthesizer output amplitude to -2
28. Press TRACE A
and DISPLAY LINE
on
spectrum analyzer.
29. Set frequency synthesizer FREQUENCY to 20 MHz. Set spectrum
analyzer
(
CENTER FREQUENCY
_)
to 20 MHz, and
(
FREQUENCY SPAN
)
to
1 MHz.
30. Repeat steps 22 through 24.
31. After completion of sweep, press DISPLAY LINE
on the
spectrum analyzer. Use the Display Line to measure the maximum
and minimum points on the trace. Record the measurements
below.
SIGNAL INPUT 2
(100
to 20 MHz)
Maximum
Minimum
Figure 2-15.
Frequency Response
Setup (100 Hz to 100
Performance Tests