3. Preliminary Display Adjustments (SN
and Below)
28. Set the oscilloscope controls as follows:
Press
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Press
29. Adjust
X POS and
Y POS to either extreme to
position the CRT beam off-screen (to prevent possible damage to
the CRT phosphor). If it is not sealed, adjust
INT GAIN
fully clockwise.
30. Adjust the spectrum analyzer’s front-panel INTENSITY control for
peak-to-peak (8 divisions) as indicated on the oscilloscope.
See Figure 3-22.
1
8 . 0 0
4 5 . 0 0 v
10.00 : dc
1 2 5 . 0 0 0
3 7 5 . 0 0 0
5 0 . 0
50.00 v
Figure 3-22.
Signal
31. Adjust
HF GAIN and
for minimum overshoot
on rise and minimum rise and fall times of the pulse waveform.
32. Use the oscilloscope
markers to measure the risetime,
falltime, and percent overshoot of the pulse waveform. Rise
and falltimes should be less than 50 ns and 90 ns respectively.
Overshoot on
the
rise should be less than 5% (approximately 0.4
divisions).
33. Set
the
spectrum analyzer’s LINE switch to STANDBY, and center
potentiometers
X POSN and
Y POSN.
34. Disconnect the oscilloscope channel 1 probe from the spectrum
analyzer. Remove
the
Display Adjustment PC board from the
slot, and reinstall the
Intensity Control Assembly.
Replace the
Section cover and cables.
35. Perform Adjustment Procedure 4,
Display Adjustment (SN
and Below).
Adjustments 3-51
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...