3. Preliminary Display Adjustments (SN
and Below)
19. Connect the oscilloscope’s channel 1 probe to
and the
channel 4 probe to
See Figure 3-18 for the location
of the test points. Connect the output of the pulse/function
generator to
(Y input) on the Display Adjustment PC board in
the
slot.
20. The Y Deflection Amplifier is identical to the X Deflection
Amplifier. Repeat steps 12 through 18 for the Y Deflection
Amplifier using
and
respectively.
Pulse Response of
21. Disconnect the oscilloscope channel 4 probe from the spectrum
Control Gate Z
analyzer. Connect the oscilloscope channel 1 probe to
Amplifier to BLANK
and connect the probe’s ground lead to chassis ground.
Input
22. On the oscilloscope, press [RECALL) (CLEAR) to perform a soft reset.
Press
CHANNEL 1
on, more preset probe , and use the
front-panel knob to set the probe to 10.00: 1. Press more .
24. Set the oscilloscope controls as follows:
Press
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.0
offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...25.0000 V
Press
time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125.000 ns
Press
level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.00000 V
Press
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Press
25. Connect the output of the Pulse/Function Generator to
(Z
input) on the Display Adjustment PC Board in the
slot. Set
the board’s switch to the down position.
Note
The pulse/function generator’s output must be terminated with 50
ohms. Use a BNC tee, a
termination, and a BNC female to SMB
female adapter. Install the
termination as close to the Display
Adjustment PC Board as possible.
26. Set the pulse/function generator’s controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..pulse
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Width(WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
27. Set the spectrum analyzer’s front-panel INTENSITY control fully
clockwise. Note the display on the oscilloscope. The pulse should
be
peak-to-peak.
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...