12. Time Base Adjustment (SN
and Below, also
and Above)
12. Time Base
Adjustment (SN
and Below,
also
and Above)
Reference
Related Performance
Test
Description
Equipment
RF Section:
10 MHz Quartz Crystal Oscillator
Center Frequency Readout Accuracy Test
The frequency of the internal 10 MHz Frequency Standard is
compared to a known frequency standard and adjusted for minimum
frequency error. This procedure does not adjust the short-term
stability or long-term stability of the 10 MHz Quartz Crystal Oscillator,
which are determined by characteristics of the particular oscillator
and the environmental and warmup conditions to which it has been
recently exposed. The spectrum analyzer must be ON continuously
(not in STANDBY) for at least 72 hours immediately prior to oscillator
adjustment to allow both the temperature and frequency of the
oscillator to stabilize.
1
Figure 3-52. Time Base Adjustment Setup
Frequency Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . .
HP
Cables:
BNC cable, 122 cm (48 in) (2 required) . . . . . . . . HP
Adjustments 3-95
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...