8. 21.4 MHz Bandwidth Filter Adjustments
10. Key in
1 MHz, and
1 MHz.
11. Press MARKER
SEARCH
],
MARKER
12. Key in
100
200
and MARKER
[
PEAK
SEARCH).
13. Adjust
LC to align markers on display. MARKER A level
should indicate 1.00 X. See Figure 3-41 for location of adjustment.
14. Repeat steps 10 through 13 until no further adjustment is
necessary.
XTAL Adjustments
15. Press MARKER
Key in
30
and
[
FREQUENCY
100
16. Press
[
REFERENCE LEVEL
)
and adjust DATA knob to set signal peak
on screen two divisions from the top graticule line.
17. Connect crystal filter bypass networks between
and
and between
and
18. Adjust
CTR to center signal on center graticule
line. Adjust
SYM for best symmetry of signal. See
Figure 3-42 for location of adjustments. If unable to adjust SYM
for satisfactory signal symmetry, increase or decrease value of
Refer to Table 3-3 for range of values.
B a n d w i d t h F i l t e r
Figure 3-42.
Location of
21.4 MHz Crystal Filter Adjustments
19. Remove crystal filter bypass network from between
and
20. Adjust
CTR to center signal on center graticule line.
Adjust
SYM for best symmetry of signal. See Figure 3-42
for location of adjustments. If unable to adjust
SYM
for satisfactory signal symmetry, increase or decrease value of
Refer to Table 3-3 for range of values.
21. Remove crystal filter bypass network from between
and
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...