18. Frequency Control Adjustments
START and STOP
Adjustments
FM SPAN Adjustment
30.
22.
23.
Key in
STEP
1023 Hz.
Adjust
LSD
for DVM indication of
V dc. If not using an HP
DVM, adjust for specified
voltage plus the DVM indication in step 20. See Figure 3-71 for
location of adjustment.
24. On the HP
press
25. Connect frequency counter to rear-panel
LO OUTPUT
connector.
26. Press
PRESET
),
then key in
STEP
CRT
annotation should indicate DACS 0.
27. Adjust
START
for frequency counter indication
of 2.050
0.002
See Figure 3-71 for location of
adjustment.
28. Key in
STEP
1023 MHz. CRT annotation should
indicate DACS 1023.
29. Adjust
STOP
STOP
for frequency counter
indication of 3.7891
See Figure 3-71 for location of
adjustment.
Press
PRESET
),
then key in
(
CENTER FREQUENCY
)
10 MHz,
[
FREQUENCY SPAN
]
20
M
HZ
.
31.
32.
Connect CAL OUTPUT to SIGNAL INPUT 2.
Adjust
FM SPAN
so that the LO Feedthrough signal is
centered on the left edge graticule and the 20 MHz CAL OUTPUT
signal is centered on the right edge graticule. See Figure 3-71 for
location of adjustment.
3-122 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...