4. Final Display Adjustments (SN
and Above)
Figure 3-32. Location of Final Display Adjustments on
3-6. Initial Adjustment Positions
Adjustment
Y GAIN
Y POSN
X GAIN
X POSN
INT GAIN
FOCUS COMP
T/B FOC
Al
T/B CTR
Al
R/L FOC
A
R/L CTR
ORTHO
INT LIM
ASTIG
Front-panel INTENSITY
Front-panel FOCUS
Front-panel ALIGN
Position
centered
centered
centered
centered
two-thirds clockwise
centered
centered
centered
centered
centered
centered
centered
fully counterclockwise
centered
fully counterclockwise
centered
centered
4. Set the spectrum analyzer’s LINE switch to ON, and wait at least
5 minutes to allow the CRT and high-voltage circuits to warm up.
5. Set the front panel INTENSITY control fully counterclockwise and
adjust
INT LIM until the display is just visable. See
Figure 3-32.
6. Set the front-panel INTENSITY control fully clockwise.
7. Adjust
X GAIN,
X POSN,
Y
GAIN, and
Y POSN for optimum centering of the
display annotation and graticule pattern.
3-62 Adjustments