25. Digital Storage Display Adjustments
Sample and Hold
13. Set LINE switch to STANDBY.
Balance Adjustments
14. Place
Line Generator on extender boards.
15. Set LINE switch to ON. Press
PRESET
].
16. Key in
(RECORDER LOWER LEFT) 0 [Hz). Press
(RECORDER UPPER RIGHT) 1028
17. Connect oscilloscope to
18. Connect
to oscilloscope External Trigger Input and
adjust oscilloscope controls for display as shown in Figure 3-91.
19. Adjust
X S&H BAL for minimum
offset between the
level of the signal inside the two pulses to the signal level outside
the two pulses. Figure 3-91 shows a properly adjusted waveform.
Figure 3-92 shows the waveform before adjustment. Refer to
Figure 3-90 for location of adjustment.
1
50.0
0 . 0 0 0 V
10.00
2 . 1 0 0 0 0
5 0 0
Figure 3-91. Sample and Hold Balance Adjustment Waveforms
1
5 0 . 0
0 . 0 0 0 v
10.00 1 dc
- 2 . 9 0 0 0 0
- 4 0 0 . 0 0 0
2 . 1 0 0 0 0
5 0 0
Figure 3-92. Waveform Before Adjustment
Adjustments