12. Option 857 Amplitude Fidelity
Frequency
1
Synthesizer Calibrated
Amplitude Amplitude
Step
0 (ref)
0
- 1 0
- 1 0
-20
-20
-30
-30
-40
-40
-50
-50
-60
-60
-70
-70
-80
-80
-90
20. Subtract the greatest negative fidelity error from the greatest
positive fidelity error for calibrated amplitude steps from -10
to -70
The results should be
21. Subtract the greatest negative fidelity error from the greatest
positive fidelity error for calibrated amplitude steps from -10
to -90
The results should be
5-2. Log Amplitude Fidelity (10
Option 857)
2
Fidelity Error
MARKER A Amplitude (Column 2 Column 1)
0 (ref)
0 (ref)
Cumulative Cumulative
Error
Error
0 to 80
0 to 90
Linear Fidelity
22. Key in analyzer settings as follows:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300
Hz
FREQUENCY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10
23. Set the frequency synthesizer for an output power level of + 10
24. Press SCALE LIN pushbutton. Press MARKER
[
PEAK
to center the signal on the display.
25.
Set
(
FREQUENCY SPAN
)
to 0 Hz and
to 1 Hz. Press (SHIFT),
(resolution bandwidth), MARKER
26. Decrease frequency synthesizer output amplitude by 10
steps,
. noting the MARKER A amplitude and recording it in column 2 of
5-3.
Option 857
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...