19. Second Converter Adjustments
Procedure
Second LO Frequency
3. Set HP
Spectrum Analyzer LINE to ON and press
and Shift Adjustments
PRESET].
Note
The second LO and pilot second LO output power is typically -35
or less. An HP
amplifier is used in steps 1 through 26 to
amplify the LO power to a
level for the counter and power
meter.
Note
1. Remove
RF Converter assembly from HP
Spectrum
Analyzer. Removal and installation procedures are contained as
a repair procedure in the RF Section of the Troubleshooting and
Repair Manual, Volume 1.
2. Position instrument on right side as shown in Figure 3-72 with the
RF Converter removed but with cables still connected.
The following adjustment tools are required to adjust the
second converter:
driver (08555-20121) and hex nut driver
(08555-20122). Place the
driver through the center hole of the
hex nut driver. Loosen the adjustment nut using the hex nut driver
while adjusting the
with the
driver. Do not over
tighten the nut on the second converter.
4. Connect the amplifier’s input to
and the power meter to
the amplifier’s output.
5. Adjust
MIXER
for maximum power meter
indication. See Figure 3-73 for location of adjustment.
R F C O N V E R T E R
2 0 5 2 5
I F B P F
Ml XER
2 0 1 7 . 6
P I L O T
I F B P F
Figure 3-73. Location of Second Converter Adjustments
6. Disconnect power meter and connect frequency counter to
amplifier’s output.
7. Adjust
LO FREQ
for frequency counter
indication of 1748.6 MHz fl.O MHz. See Figure 3-73 for location
of adjustment.
8. Disconnect frequency counter and reconnect power meter to
amplifier’s output.
24 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...