14. 249 MHz Phase Lock Oscillator Adjustments
14. 249 MHz Phase
Lock Oscillator
Adjustments
Reference
Description
Equipment
RF Section:
249 MHz Phase Lock Oscillator
Two center frequencies are chosen: one which will tune the 249 MHz
Oscillator to its low-end frequency and one which will tune the 249
MHz Oscillator to the high-end frequency. The voltage is monitored
with a DVM at the output of the oscillator, and the oscillator
frequency is adjusted to produce the proper dc voltage output for
each frequency (low-end and high-end). Next, the 500
Trap is
adjusted to null the 500
sidebands using the sixth harmonic of the
249 MHz signal.
ANALYZER
DIGITAL VOLTMETER
SPECTRUM ANALYZER
J
Figure 3-59. 249 MHz Phase Lock Oscillator Adjustments Setup
Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . .
HP
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . HP
Tee, SMB Male . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 1250-0670
Adapters:
Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1250
Cables:
BNC cable, 122 cm (48 in) (2 required) . . . . . . . . . . . .
BNC to SMB Snap-On cable (2 required) . . . . . . 85680-60093
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...