8. Frequency Response Test
100
32.
33.
34.
35.
36.
37.
38.
39.
40.
41.
Press
PRESET
)
on the spectrum analyzer. Activate SIGNAL
INPUT 1.
Key in the following spectrum analyzer settings:
START
. . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
. . . . . . . . . 100
Connect equipment as shown in Figure 2-15 with function
generator to SIGNAL INPUT 1.
Set the function generator controls as follows:
LINE . . . . . . . . . . . . . . . . . . .. . . . . . .. . . . . . . . . . . . . . . . . . . . ON
RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 K
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL (button in)
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 V
AMPLITUDE VERNIER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . midrange
SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
TRIGGER PHASE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FREE RUN
MODULATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . all out
MODULATION RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I
M O D U L A T I O N R A N G E H z V E R N I E R . . . . . . . f u l l y C C W
MODULATION SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
Percent Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CW
Adjust function generator FREQUENCY to place signal between
the last two graticule lines (right side) on the signal analyzer
display.
Adjust AMPLITUDE VERNIER on the function generator until
the peak of the signal is at the reference graticule line on the
spectrum analyzer display.
Press LOG
1
on the spectrum analyzer. Press
DISPLAY LINE
and set the Display Line to the level
recorded for 100
in step 25.
Adjust function generator AMPLITUDE VERNIER to place peak of
signal at the Display Line.
Adjust FREQUENCY on the function generator to position the
signal trace at the right edge of the spectrum analyzer display
(last graticule line).
Press MODULATION SWP on the function generator and allow the
function generator to make at least two complete sweeps. Press
TRACE A [MAX HOLD). Allow the function generator to make one
complete sweep. After completion of the sweep, press TRACE A
Performance Tests
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...