24. Track and Hold
Adjustments
Reference
Track and Hold
Description
The CAL OUTPUT signal is connected to the RF INPUT. The
instrument is placed in zero frequency span to produce a dc level
output from the IF-Video section and this dc level is regulated by
adjusting the reference level. The Offsets and Gains on the Track and
Hold assembly are adjusted for proper levels using a DVM.
SPECTRUM
ANALYZER
DIGITAL
Figure 3-87. Track and Hold Adjustments Setup
Equipment
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Procedure
1. Place instrument upright as shown in Figure 3-87 with top and
Digital Storage covers removed.
2. Set LINE switch to ON and press
(
INSTR PRESET
].
3. Connect CAL OUTPUT to RF INPUT.
4. Connect DVM to
and ground to
5.
Key in
FREQUENCY
)
20
MHz,
SPAN
]
0 Hz.
6. Disconnect cable 7 (violet) from
7. Short
to
or use an SMB snap-on short to
DVM indication should be 0.000
V
8. Key in
TRACE A
(
CLEAR
-
WRITE
],
MARKER
MARKER
SWEEP
TRACE A
9. Adjust
(T/H) OFS until MARKER
level indication as
indicated by CRT annotation flickers back and forth between .OO
and
See Figure 3-88 for location of adjustment.
3-142 Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...