23. Analog-To-Digital Converter Adjustments
23.
Analog-To-Digital
Converter
Adjustments
Reference
Description
Equipment
Analog-to-Digital Converter
The Analog-to-Digital Ramp Converter is adjusted at zero and
full-scale by injecting a 0 V dc input and + 10 V dc input and adjusting
the OFFS and GAIN controls until the ramp output at
toggles high to low. This sets the horizontal end points for the CRT
trace display; when the sweep ramp input is at 0 V dc (the left
graticule edge), trace position 1 is set, and when the sweep ramp
input is at + 10 V
(the right graticule edge), trace position 1000 is
set.
This procedure requires a + 10 V dc source which is stable and
noise-free. A simple supply circuit which can be built with common
components is illustrated in Figure 3-93. If these components are
unavailable, the alternate procedure provided below (using only the
digital voltmeter) can then be used.
DIGITAL VOLTKTER
. -
1’
HP
OSCILLOSCOPE
Figure 3-85. Analog-To-Digital Converter Adjustments Setup
Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 54501
Digital Voltmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
Low-Noise DC Supply (Optional) . . . . . . . . . . . . . . . . . . . . . See Figure 3-93
Adjustments 3-139
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...