13. 20 MHz Reference Adjustments
13. 20 MHz
Reference
Adjustments
Reference
Related Performance
Description
Equipment
RF Section:
20 MHz Reference
Calibrator Amplitude Accuracy Test
The 20 MHz output is peaked and amplitude checked for proper level.
The INTERNAL REFERENCE output level is then checked for proper
output level as compared to input from
Time Base. Finally, the
COMB DRIVE and CAL OUTPUT are adjusted for proper power levels.
SYNTHESIZER LEVEL
SPECTRUM ANALYZER
POWER METER
DIGITAL VOLTMETER
(Devise Under
Figure 3-56. 20
Reference Adjustments Setup
Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . .
HP
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . HP
Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . HP
Power Meter
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Digitizing Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . .
HP
Divider Probe . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
Adapters:
Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . 1250-0780
Type N (f) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1474
Cables:
BNC to SMB cable Snap-On (2 required) . . . . . . 85680-60093
Adjustments
Summary of Contents for 8568B
Page 13: ...6 7 IF Section Bottom View 6 10 Contents 7 ...
Page 19: ......
Page 37: ......
Page 42: ......
Page 48: ......
Page 53: ......
Page 61: ......
Page 90: ......
Page 106: ...Test 13 Average Noise Level Test 2 84 Performance Tests ...
Page 162: ......
Page 224: ......
Page 268: ......
Page 270: ......
Page 272: ......
Page 278: ......
Page 312: ...AlVl Figure 6 6 IF Section Front View w3 w9 Major Assembly and Component locations 6 9 ...