OPERATI ONAL VERI FICAT ION
MOD E L 3562
2-1 4
Procedure
A. Connect the test equipment as shown in figure 2-5. Refer to "Initial Equ ipment Setu p,"
section 2-6, for unspecified parameters.
B. Set the test instru ments initial ly as fol lows:
Frequency Synthesizer
Frequency
Ampl itude
Function
99 kHz
1 Vrms
Sine Wave
C. Press the H P 3562A keys as follows:
PRESET
CAl
RAN G E
FREQ
AVG
START
x
RESET
SI NGLE
CAL
o
dBVrms
CE NTE R FREQ
2
STABLE
99
kHz
ENTER
D. Record the X marker read ing as the measured value on the Operational Verification
Test Record.
I f Test Fails Check:
Adjustments
Section I I I
Troubleshooting
Section VI I
20.48 MHz Reference Adj ustment
A31 Trigger Board
Summary of Contents for 3562A
Page 2: ......
Page 6: ......
Page 16: ...GEN ERAL INFORMATION MODEL 3562 T bJe 1 3 Specifications cont 1 10 ...
Page 20: ......
Page 24: ......
Page 126: ......
Page 128: ......
Page 150: ......
Page 152: ......
Page 160: ......
Page 196: ......
Page 198: ......
Page 206: ......
Page 207: ...MODEL 3562A CR Cl ...
Page 209: ...MODE L 3562A Cl ...
Page 211: ... ...
Page 213: ...r A1 a pQWERSuP PLY I 03562 66518 REV A REV 8 8MPOUT 58 58 FRONT REAR P ANEL ii O N ...
Page 214: ... C401 8S1 15ISI t 1 J400 ...
Page 217: ... ...
Page 220: ......
Page 221: ... ...
Page 224: ...A3 CQVLCLI A3 ...
Page 231: ...S V InO 3J nOS N I l3 3 1 1 1X3 NI 31dWVS lX3 H l 1 3 NNVH I 0 Ioe J ...
Page 237: ...Sample Clock SAMP DS DATA oe _ _ _ _ _ BUS NOTE FULL SPA BASi BAI MODE ...
Page 246: ......
Page 259: ...WRITEL A22L l cc E Vl 8 MHz A23L ASL Inverting A1L A21 L D rivers r ...
Page 305: ...c c I O Sequencer S r J Sequence Decoder ...
Page 320: ......
Page 324: ...EXT TRIGGER TRIG 1 TRIG 2 CALTRIG CNTLD COMI W CNTCLK LDTRGL RESETL ...
Page 327: ... RH r I N EXT BUFFER SAMPLE _ _ _ IN ...
Page 450: ......
Page 488: ......
Page 492: ......
Page 536: ......
Page 552: ......
Page 570: ......