ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
82
000000-2071-464
10/2014 V_01
Loops and Repetitions are set (in
Show all
mode)
in the
Loops and Repetition
panel. Use the arrow
to open the panel (Fig. 94). The number in the
Loops
column states how often a sequence of
Blocks defined by the number of the Block in
Start
and
End
is run. One loop/repetition is set as
default and reflects the minimum of running the
experiment once. Set the Loops value to 2 to
perform the defined sequence twice. Repeat one
Block by choosing the same number for Start and
End. A yellow triangle shows up if a sequence
cannot be performed for example if the end block
is before the start block. Whenever a block being
part of a sequence is reached, the defined number
of loops is performed. Thus outer loops also contribute to the number of total runs of an inner loop. Add
a new row of loop settings by clicking the + button. Highlighted rows are removed by clicking the waste
bin button. Rows of settings need to be checked in the
Enabled
column to be executed.
Use the
Export…
button to store a complete Multi Block experiment or selected Blocks under a
Name
of
choice typed into the editing field. Within the
Export Experiment Blocks
dialogue you can
Select All
or
Deselect All
blocks using the corresponding buttons. Single blocks are selected/deselected with a mouse
click.
Export
stores the selected block(s) under the chosen name.
Use the
Import…
button to load a Multi Block experiment or selected Blocks of an Experiment. Within
the
Import Experiment Blocks
dialogue chose a name from the
Experiments
list with mouse click.
Select All
or
Deselect All
blocks using the corresponding buttons. Single blocks are selected/deselected
with a mouse click. Delete removes the highlighted name/experiment from the list.
Import
adds the
selected block(s) to the already existing block(s) of the Experiment Designer.
Pay attention to the following guidelines for the Experiment Designer
−
Each Block and looped Block is (saved as) a separate image file
−
Auto Save settings apply to the complete Multi Block experiment
−
Image files are automatically named consecutively keeping the number for the image constant for a
given Multi Block Experiment. The Block number always starts with 1 and the number is
consecutive within the experiment.
−
Use the Macro "ConcatenateImages70.lvb" to combine still open images into one consecutive time
series. See Chapter on Macros for further explanations.
−
Use the Macro "ModifySeriesZEN.lvb" to combine already already saved images into consecutive
time series iamges. See Chapter on Macros for further explanations.
−
Set desired stage position(s) for each new block separately using the Positions Tool. For multi-
position experiments in combination with Z-Stack acquisition the Z coordinate from the Position list
is referring to the center position of the Z-Stack.
−
Bleaching experiments are restricted to one position per Block. If more than one position is defined
for a Block, Bleaching is disabled.
−
Positions and Regions are not saved with the experiment, only with the image. Positions cannot be
reused from the resulting image but have to be stored as separate position list(s). See also section
5.2.20.1 on using position lists.
Fig. 94
Loops and Repetitions panel of the
Experiment Designer tool
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......