ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
398
000000-2071-464
10/2014 V_01
5.5.1.2
Collimators Section Panel
The
Collimators Section
panel has the following options (Fig. 555):
Name
drop down menu
Selection of the collimator (IR / VIS or UV / VIS) to be adjusted via
the
drop down menu
.
Description
field
Display of the relevant active Collimator
Position [mm]
field
Setting of collimator position using the slider or arrow buttons; the
display to the left of the slider indicates the current position
Status display for setting procedure: green for ready and red for busy.
Current Position
button
Stores the current collimator position.
Stored Pos
button
Sets the collimator to the stored value.
Adjustment Move to Opt Pos
button
Starts the automatic collimator adjustment. Available for some
objectives.
Fig. 555
Collimator section panel
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......